BCI 3304
The Autocorr® Pulse Oximeter with SAC (Serial Autocorrelation) Technology uses refined hardware and unique
patented software to extract SpO2, pulse strength, and pulse rate data in a new way. This cost-effective compact monitor incorporates clinical, home and sleep
study functions as standard features, and a data log or trend printout of stored information is available when used in conjunction with a printer or computer.
Uses refined noise-reducing hardware and a unique patented software algorithm. Instead of relying on a voltage threshold to determine a pulse, it analyzes a
digitized signal at a rate of 120 times per second, and compares it to previous pulse data. These pulses are constantly stored and updated, and used as dynamic
templates to determine the validity or invalidity for newly-encountered pulsatile signals. Patented Serial Autocorrelation (SAC) technology provides
reliable monitoring.
Specifications:
- SpO2:
- Range: 0-100%
- Accuracy: ±2%
@70-100% ±3% @50-69% Sp02 - Alarm Range
High=50-100% & OFF (1% steps); Low=50-99% & OFF (1 % steps) - Averaging 4, 8, or 16
second average
- Pulse Rate:
- Range; 30-254 bpm
- Accuracy: ±2 bpm or
±2% (whichever is greater) @ 30-254 bpm - Alarm Range
High=5-250 bpm & OFF; Low=5-250 bpm & OFF - Averaging: 8 or 16
second average
- Dimensions: 3.24"Hx8.5"Wx5.5"D
(216mm x 82mm x 140mm) - AC Power Requirements: Two
voltage ranges selected by external switch (90-125VAC 50/60Hz; 200-240VAC
50/60Hz) - Battery: Standard
lead-acid battery: Internal rechargeable; 4.5 hour continuous use (new) - Printer Output: SpO2 and
pulse rate can be printed every five seconds (data log). Data saved every
four to thirty seconds can be printed (trend)